CTA Upper Extremity Run-Off w/IVcon | IMG12132 | |
Machine Model: | Canon Genesis 640 |
Generated:
2021-04-20 20:31:58
|
|
|
Special Instructions: Supine, Extremity of interest is placed by patient’s side. Other arm should be placed over patient’s head. PIV should not be in arm of concern. | |||||
Scanogram: AP/Lateral Landmark: Apices | |||||
BT | Bolus Tracking | A | Arterial | ||
Scan Start | Aortic Arch | Thoracic Inlet | |||
Scan Stop | Aortic Arch | Below Hands | |||
Patient Position | Supine | Supine | |||
Scan Direction | NA | Craniocaudal | |||
Scan Mode | S&V | Helical | |||
Scan Range If Volume | NA | NA | |||
Scan Delay | 20 seconds | NA | |||
Bolus Tracking | Aorta 200 HU | Aorta 200 HU | |||
Reconstruction 1 | Thickness x Spacing | 1 mm x .8 mm | |||
Algorithm | CTA Body | ||||
FOV | 40 cm | ||||
Reconstruction 2 | Thickness x Spacing | 1 mm x .8 mm | |||
Algorithm | CTA Body | ||||
FOV | Size/Cone to Extremity | ||||
Reconstruction 3 | Thickness x Spacing | MIP 10 x 5 | |||
Algorithm | CTA Body | ||||
FOV | Coronal + Sagittal | ||||
FC | 1 | 18 | |||
kVp | 120 kVp | Auto* kVp | |||
Scan FOV | Patient Largest + 4CM | Patient Largest + 4CM | |||
Pitch | Standard | Standard | |||
mA | 50 mA | modulated mA | |||
Rotation Time | .5 sec | .5 sec | |||
Detector Rows | 1 mm x 4 mm | 0.5 mm x 100 mm | |||
Indications: No indications found. . |