CTA Cardiac Post Graft IVcon wo/w | IMG12121 | |
Machine Model: | Canon Aquilion ONE 320, Canon Genesis 640, Canon Vision 640 |
Generated:
2024-08-16 13:27:38
|
|
|
Special Instructions: Supine, Feet first into gantry, Heart isocenter. 80ml contrast. Refer to bottom of page for further details. | |||||||
Scanogram: AP/Lateral Landmark: Above Shoulders | |||||||
A | Calcium Score | BT | Bolus Tracking | B | Gated Cardiac | ||
Scan Start | Top of Heart | Carina | Above Subclavian Artery | ||||
Scan Stop | Bottom of Heart | Carina | Bottom of Heart | ||||
Patient Position | Supine | Supine | Supine | ||||
Scan Direction | NA | NA | NA | ||||
Scan Mode | Volume | S&V | Helical | ||||
Scan Range If Volume | 160 mm | NA | NA | ||||
Scan Delay | NA | 10 seconds | NA | ||||
Bolus Tracking | NA | Aorta 180 HU | Aorta 180 HU | ||||
Reconstruction 1 | Thickness x Spacing | 0.5 mm x 0.25 mm | .5 mm x .3 mm | ||||
Algorithm | Calcium Score | Cardiac CTA | |||||
FOV | 22 cm | 32 cm | |||||
Cardiac Phase |
Determined by Heart Rate |
75% |
|||||
Reconstruction 2 | Thickness x Spacing | 3 mm x 3 mm | .5 mm x .3 mm | ||||
Algorithm | Calcium Score | Cardiac CTA | |||||
FOV | 22 cm | 32 cm | |||||
Cardiac Phase |
Determined by Heart Rate |
Best Phase |
|||||
FC | 12 | 1 | 3 | ||||
kVp | 120 kVp | 120 kVp | 100 kVp | ||||
Scan FOV | 22 cm | 22 cm | 32 cm | ||||
Pitch | NA | NA | NA | ||||
mA | Sure Exp Calcium Score | 50 mA | Sure Exp CTA Standard | ||||
Rotation Time | 0.275 sec | 0.275 sec | 0.275 sec | ||||
Detector Rows | 0.5 mm x 160 mm | 0.5 mm x 4 mm | 0.5 mm x 160 mm | ||||
Indications: chest pain , coronary artery disease , need for definition of anatomy prior to EP ablation , need for definition of anatomy prior to surgery , patency of bypass grafts . |