CT Chest and Abd Pelvis Ped w/IVcon | IMG14096 | |
Machine Model: | Canon AQ Prime |
Generated:
2021-04-20 10:32:07
|
|
|
Special Instructions: Above lung apices to lesser trochanter | |||
Scanogram: AP/Lateral Landmark: Apices | |||
A | Venous | ||
Scan Start | Apices | ||
Scan Stop | Lesser Trochanter | ||
Patient Position | Supine | ||
Scan Direction | Craniocaudal | ||
Scan Mode | Helical | ||
Scan Range If Volume | NA | ||
Scan Delay | 60 seconds | ||
Reconstruction 1 | Thickness x Spacing | 1 mm x 1 mm | |
Algorithm | Ped Body | ||
FOV | 24 cm | ||
Reconstruction 2 | Thickness x Spacing | 1 mm x 1 mm | |
Algorithm | Lung STD | ||
FOV | RIB to RIB | ||
FC | 18 | ||
kVp | See Chart Below Based on Weight | ||
Scan FOV | Patient largest + 2CM | ||
Pitch | Standard | ||
mA | modulated mA | ||
Rotation Time | .5 sec | ||
Detector Rows | 0.5 mm x 64 mm | ||
Indications: FUO (fever of unknown origin) in immunocompromised host , PTLPD - include neck , aplastic anemia and neutropenia , infection , lymphoma - include neck , malignant tumors and chemotherapy f/u , opsiclonus/myoclonus , trauma , vasculopathy . |
15KG-30KG | 0.35 Rotation Time |
31KG - 61KG+ | 0.5 Rotation Time |