26133
26142
CT Angio Abd Pelvis IVcon IMG12113
Machine Model: Canon AQ Prime, Canon Aquilion ONE 320, Canon Vision 640
Generated:
2021-05-10 10:54:48
IV Contrast 4 ml/sec using 20 g PIV Minimum See IV Guidelines
for other access lines.
IV Saline NA
Oral Contrast NA
Water NA
Contrast amount determined by weight and renal function.
Special Instructions: Dome of Diaphragm to Lesser Trochanter
Scanogram: AP/Lateral
Landmark: Dome of Diaphram
BT Bolus Tracking A Arterial
Scan Start Dome of Diaphram Dome of Diaphram
Scan Stop Dome of Diaphram Lesser Trochanter
Patient Position Supine Supine
Scan Direction NA Craniocaudal
Scan Mode S&V Helical
Scan Range If Volume NA NA
Scan Delay 20 seconds NA
Bolus Tracking Aorta 200 HU Aorta 200 HU
Reconstruction 1 Thickness x Spacing 1 mm x .8 mm
Algorithm Body STD
FOV 40 cm
Reconstruction 2 Thickness x Spacing MIP 10 x 5
Algorithm Body STD
FOV Coronal + Sagittal
FC 3 18
kVp 120 kVp 120 kVp
Scan FOV Patient Largest + 4CM Patient Largest + 4CM
Pitch Standard Standard
mA 50 mA modulated mA
Rotation Time .5 sec .5 sec
Detector Rows 1 mm x 4 mm 0.5 mm x 100 mm
Indications: abdominal aortic aneurysm , abdominal pain , aortic aneurysm , aortic anomaly , aortic dissection , aortic occlusion , aortic stenosis , body trauma , post-op , presurgical planning , renal artery AVM / arteriovenous malformation , renal artery aneurysm , renal artery arteriovenous fistula , renal artery stenosis , traumatic aortic injury , traumatic injuries to arteries and veins , unspecified aorta pathology , vascular interventional planning .