19195
19202
CT Maxface Dtl and Limited Neck w/IVcon IMG14059
Machine Model: Canon AQ Prime
Generated:
2021-05-10 13:23:51
IV Contrast 0.8 ml/sec using 24 g PIV Minimum See IV Guidelines
for other access lines.
IV Saline NA
Oral Contrast NA
Water NA
Contrast amount determined by weight and renal function.
Special Instructions: IOML perpendicular to table top, No Angle, Include tip of nose *stealth*,Neck angle with vertebral disc space
Scanogram: Lateral / AP
Landmark: Below Clavicles
A Venous B Venous
Scan Start Below Mentum Mid Oral
Scan Stop Above Frontal Sinus C4-C5 Disc Space
Patient Position Supine Supine
Scan Direction caudal/cranial Craniocaudal
Scan Mode Helical Helical
Scan Range If Volume NA NA
Scan Delay 60 seconds NA
Reconstruction 1 Thickness x Spacing 1 mm x .8 mm 1 mm x 1 mm
Algorithm Body STD Body STD
FOV 22 cm 18 cm
Reconstruction 2 Thickness x Spacing .5 mm x .3 mm
Algorithm Bone
FOV 22 cm
FC 8 8
kVp 120 kVp 120 kVp
Scan FOV 22 cm 18 cm
Pitch Standard Standard
mA 250 mA modulated mA
Rotation Time .5 sec .5 sec
Detector Rows 0.5 mm x 80 mm 0.5 mm x 80 mm
Indications: atypical facial pain , benign and malignant tumors , benign masses and low grade malignancy dental origin , esthesioneuroblastoma , facial abscess , first branchial cleft cyst , inflammatory disease limited perimandibular infect , inflammatory disease of submandibular gland , known or very likely benign mass , nasopharynx central skull base known or very likely benign mass (e.g., parapharyngeal space benign mixed tumor , oral infection , preauricular benign mass , ranula , tongue weakness and atrophy , trigeminal neuralgia .