CT Mandible No IVcon | IMG14054 | |
Machine Model: | Canon AQ 16 |
Generated:
2021-05-10 13:19:30
|
|
|
Special Instructions: IOML perpendicular to table top | |||
Scanogram: Lateral / AP Landmark: Hyoid | |||
A | Non-Contrast | ||
Scan Start | Below Mentum | ||
Scan Stop | Above Frontal Sinus | ||
Patient Position | Supine | ||
Scan Direction | caudal/cranial | ||
Scan Mode | Helical | ||
Scan Range If Volume | NA | ||
Scan Delay | NA | ||
Reconstruction 1 | Thickness x Spacing | 1 mm x .8 mm | |
Algorithm | Body STD | ||
FOV | 25 cm | ||
Reconstruction 2 | Thickness x Spacing | .5 mm x .3 mm | |
Algorithm | Bone | ||
FOV | 18 cm | ||
FC | 23 | ||
kVp | 120 kVp | ||
Scan FOV | 22 cm | ||
Pitch | Detail | ||
mA | 250 mA | ||
Rotation Time | .5 sec | ||
Detector Rows | 0.5 mm x 16 mm | ||
Indications: TMJ / temporomandibular joint , congenital or growth related acquired deformities , facial bones trauma , mandible trauma , metabolic disease . |