CT Orbits w/IVcon IMG14072
Machine Model: Canon Aquilion ONE 320
Generated:
2021-05-10 13:41:45
IV Contrast 0.8 ml/sec using 24 g PIV Minimum See IV Guidelines
for other access lines.
IV Saline NA
Oral Contrast NA
Water NA
Contrast amount determined by weight and renal function.
Special Instructions: IOML perpendicular to table top, No Angle, Include tip of nose *stealth*
Scanogram: Lateral / AP
Landmark: Hyoid
A Venous
Scan Start Hard Palate
Scan Stop Above Frontal Sinus
Patient Position Supine
Scan Direction caudal/cranial
Scan Mode Helical
Scan Range If Volume NA
Scan Delay 90 seconds
Reconstruction 1 Thickness x Spacing 1 mm x .8 mm
Algorithm Body STD
FOV 14 cm
Reconstruction 2 Thickness x Spacing .5 mm x .3 mm
Algorithm Bone
FOV 14 cm
FC 8
kVp 120 kVp
Scan FOV 24 cm
Pitch Detail
mA 250 mA
Rotation Time .5 sec
Detector Rows 0.5 mm x 64 mm
Indications: exophthalmos , inflammatory disease of orbit , ocular masses , optic nerve and more posterior visual pathways , orbital masses , orbital varix , retinoblastoma follow-up study .