CT BRW/Radiosurgery/CRW w/IVcon IMG14024
Machine Model: Canon Vision 640
Generated:
2021-05-10 11:00:34
IV Contrast 0.8 ml/sec using 24 g PIV Minimum See IV Guidelines
for other access lines.
IV Saline NA
Oral Contrast NA
Water NA
Contrast amount determined by weight and renal function.
Special Instructions: IOML perpendicular to table (ring bolted to table), No Angle, 350 FOV, 1x1, Bottom ring through top ring
Scanogram: Lateral / AP
Landmark: Hyoid
A Venous
Scan Start Bottom BRW Ring
Scan Stop Top BRW Ring
Patient Position Supine
Scan Direction caudal/cranial
Scan Mode Helical
Scan Range If Volume NA
Scan Delay 2 minutes
Reconstruction 1 Thickness x Spacing 1 mm x 1 mm
Algorithm Brain
FOV 35 cm
FC 64
kVp 120 kVp
Scan FOV 35 cm
Pitch Detail
mA 240 mA
Rotation Time .75 sec
Detector Rows 0.5 mm x 64 mm
Indications: No indications found. .