CT BRW/Radiosurgery/CRW IVcon wo/w IMG14025
Machine Model: Canon Vision 640
Generated:
2021-05-10 10:59:47
IV Contrast 0.8 ml/sec using 24 g PIV Minimum See IV Guidelines
for other access lines.
IV Saline NA
Oral Contrast NA
Water NA
Contrast amount determined by weight and renal function.
Special Instructions: IOML perpendicular to table (ring bolted to table), No Angle, 350 FOV, 1x1, Bottom ring through top ring
Scanogram: Lateral / AP
Landmark: Hyoid
A Non-Contrast B Venous
Scan Start Bottom BRW Ring Bottom BRW Ring
Scan Stop Top BRW Ring Top BRW Ring
Patient Position Supine Supine
Scan Direction caudal/cranial caudal/cranial
Scan Mode Helical Helical
Scan Range If Volume NA NA
Scan Delay NA 2 minutes
Reconstruction 1 Thickness x Spacing 1 mm x 1 mm 1 mm x 1 mm
Algorithm Brain Brain
FOV 35 cm 35 cm
FC 64 64
kVp 120 kVp 120 kVp
Scan FOV 35 cm 35 cm
Pitch Detail Detail
mA 240 mA 240 mA
Rotation Time .75 sec .75 sec
Detector Rows 0.5 mm x 64 mm 0.5 mm x 64 mm
Indications: No indications found. .