CTA Head and Neck IVcon wo/w | IMG12134 | |
Machine Model: | Canon Genesis 640 |
Generated:
2021-04-20 20:30:36
|
|
|
Special Instructions: IOML perpendicular to table top | |||||||||||
Scanogram: Lateral / AP Landmark: Carina | |||||||||||
A | Non-Contrast | BT | Bolus Tracking | B | Arterial | C | Arterial | D | Venous | ||
Scan Start | Skull Base | Hyoid | Skull Base | Sella | Skull Base | ||||||
Scan Stop | Skull Vertex | Hyoid | Skull Vertex | Carina | Skull Vertex | ||||||
Patient Position | Supine | Supine | Supine | Supine | Supine | ||||||
Scan Direction | caudal/cranial | NA | caudal/cranial | Craniocaudal | caudal/cranial | ||||||
Scan Mode | Helical | S&V | Helical | Helical | Helical | ||||||
Scan Range If Volume | NA | NA | NA | NA | NA | ||||||
Scan Delay | NA | 10 seconds | NA | NA | 2 minutes | ||||||
Bolus Tracking | NA | Bolus Track Carotids Manually | Bolus Track Carotids Manually | NA | NA | ||||||
Reconstruction 1 | Thickness x Spacing | 1 mm x 1 mm | .5 mm x .3 mm | 1 mm x .8 mm | 1 mm x 1 mm | ||||||
Algorithm | Brain | CTA Brain | CTA Neck | Brain | |||||||
FOV | 22 cm | 22 cm | 24 cm | 22 cm | |||||||
Reconstruction 2 | Thickness x Spacing | .5 mm x .3 mm | MIP 10 x 5 | MIP 10 x 5 | |||||||
Algorithm | CTA Brain | CTA Brain | CTA Neck | ||||||||
FOV | 22 cm | Axial + Cor + Sag | Coronal + Obliques | ||||||||
FC | 64 | 1 | 43 | 43 | 64 | ||||||
kVp | 120 kVp | 120 kVp | 120 kVp | 120 kVp | 120 kVp | ||||||
Scan FOV | 24 cm | 24 cm | 24 cm | 24 cm | 24 cm | ||||||
Pitch | Detail | NA | Detail | Detail | Detail | ||||||
mA | 400 mA | 50 mA | 400 mA | Quality Modulated mA | 400 mA | ||||||
Rotation Time | .5 sec | .5 sec | .5 sec | .5 sec | .75 sec | ||||||
Detector Rows | 0.5 mm x 80 mm | 0.5 mm x 8 mm | 0.5 mm x 80 mm | 0.5 mm x 80 mm | 0.5 mm x 80 mm | ||||||
Indications: Horners Syndrome , aneurysm , stenosis , subarachnoid hemorrhage , vascular abnormality , vascular bypass surgery , vertebrobasilar insufficiency . |