| CT Head Craniosynostosis No IVcon | IMG15377 | |
| Machine Model: | Canon AQ Prime |
Generated:
2021-04-20 15:52:54
|
|
|
|||||||||||||||
| Special Instructions: IOML perpendicular to table top | |||
| Scanogram: Lateral / AP Landmark: Hyoid | |||
| A | Non-Contrast | ||
| Scan Start | Skull Base | ||
| Scan Stop | Skull Vertex | ||
| Patient Position | Supine | ||
| Scan Direction | caudal/cranial | ||
| Scan Mode | Helical | ||
| Scan Range If Volume | NA | ||
| Scan Delay | NA | ||
| Reconstruction 1 | Thickness x Spacing | 1 mm x .8 mm | |
| Algorithm | Ped Brain | ||
| FOV | 20 cm | ||
| Reconstruction 2 | Thickness x Spacing | .5 mm x .3 mm | |
| Algorithm | Bone | ||
| FOV | 20 cm | ||
| Reconstruction 3 | Thickness x Spacing | 3D x | |
| Algorithm | SPIN | ||
| FOV | TUMBLE | ||
| FC | 49 | ||
| kVp | 120 kVp | ||
| Scan FOV | 20 cm | ||
| Pitch | Detail | ||
| mA | 10 mA | ||
| Rotation Time | .5 sec | ||
| Detector Rows | 0.5 mm x 40 mm | ||
| Indications: No indications found. . | |||