13846
13857
CT Abdomen Pelvis Liver 4PH IVcon wo/w IMG14297
Machine Model: Canon Genesis 640
Generated:
2021-05-10 10:49:20
IV Contrast 4 ml/sec using 20 g PIV Minimum See IV Guidelines
for other access lines.
IV Saline NA
Oral Contrast NA
Water 2-8 oz or 472 ml
Contrast amount determined by weight and renal function.
Special Instructions: Dome of Diaphragm to Lesser Trochanter
Scanogram: AP/Lateral
Landmark: Dome of Diaphram
A Non-Contrast BT Bolus Tracking B Arterial C Venous D Delay
Scan Start Dome of Diaphram Dome of Diaphram Dome of Diaphram Dome of Diaphram Dome of Diaphram
Scan Stop 2 cm below Iliac Crest Dome of Diaphram 2 cm below Iliac Crest Lesser Trochanter 2 cm below Iliac Crest
Patient Position Supine Supine Supine Supine Supine
Scan Direction Craniocaudal NA Craniocaudal Craniocaudal Craniocaudal
Scan Mode Helical S&V Helical Helical Helical
Scan Range If Volume NA NA NA NA NA
Scan Delay NA 20 seconds 20 seconds 70 seconds 3 minutes
Bolus Tracking NA Above diaphragm
200 HU
NA NA NA
Reconstruction 1 Thickness x Spacing 1 mm x 1 mm 1 mm x 1 mm 1 mm x 1 mm 1 mm x 1 mm
Algorithm Body STD Body STD Body STD Body STD
FOV 40 cm 40 cm 40 cm 40 cm
FC 18 3 18 18 18
kVp Auto* kVp 120 kVp Auto* kVp Auto* kVp Auto* kVp
Scan FOV Patient Largest + 4CM Patient Largest + 4CM Patient Largest + 4CM Patient Largest + 4CM Patient Largest + 4CM
Pitch Standard NA Standard Standard Standard
mA modulated mA 50 mA modulated mA modulated mA modulated mA
Rotation Time .5 sec .5 sec .5 sec .5 sec .5 sec
Detector Rows 0.5 mm x 80 mm 1 mm x 4 mm 0.5 mm x 80 mm 0.5 mm x 80 mm 0.5 mm x 80 mm
Indications: liver post embolization , transcatheter arterial chemoembolization .