CT Abdomen Pelvis Liver 4PH IVcon wo/w | IMG14297 | |
Machine Model: | Canon Genesis 640 |
Generated:
2021-05-10 10:49:20
|
|
|
Special Instructions: Dome of Diaphragm to Lesser Trochanter | |||||||||||
Scanogram: AP/Lateral Landmark: Dome of Diaphram | |||||||||||
A | Non-Contrast | BT | Bolus Tracking | B | Arterial | C | Venous | D | Delay | ||
Scan Start | Dome of Diaphram | Dome of Diaphram | Dome of Diaphram | Dome of Diaphram | Dome of Diaphram | ||||||
Scan Stop | 2 cm below Iliac Crest | Dome of Diaphram | 2 cm below Iliac Crest | Lesser Trochanter | 2 cm below Iliac Crest | ||||||
Patient Position | Supine | Supine | Supine | Supine | Supine | ||||||
Scan Direction | Craniocaudal | NA | Craniocaudal | Craniocaudal | Craniocaudal | ||||||
Scan Mode | Helical | S&V | Helical | Helical | Helical | ||||||
Scan Range If Volume | NA | NA | NA | NA | NA | ||||||
Scan Delay | NA | 20 seconds | 20 seconds | 70 seconds | 3 minutes | ||||||
Bolus Tracking | NA |
Above diaphragm 200 HU |
NA | NA | NA | ||||||
Reconstruction 1 | Thickness x Spacing | 1 mm x 1 mm | 1 mm x 1 mm | 1 mm x 1 mm | 1 mm x 1 mm | ||||||
Algorithm | Body STD | Body STD | Body STD | Body STD | |||||||
FOV | 40 cm | 40 cm | 40 cm | 40 cm | |||||||
FC | 18 | 3 | 18 | 18 | 18 | ||||||
kVp | Auto* kVp | 120 kVp | Auto* kVp | Auto* kVp | Auto* kVp | ||||||
Scan FOV | Patient Largest + 4CM | Patient Largest + 4CM | Patient Largest + 4CM | Patient Largest + 4CM | Patient Largest + 4CM | ||||||
Pitch | Standard | NA | Standard | Standard | Standard | ||||||
mA | modulated mA | 50 mA | modulated mA | modulated mA | modulated mA | ||||||
Rotation Time | .5 sec | .5 sec | .5 sec | .5 sec | .5 sec | ||||||
Detector Rows | 0.5 mm x 80 mm | 1 mm x 4 mm | 0.5 mm x 80 mm | 0.5 mm x 80 mm | 0.5 mm x 80 mm | ||||||