12481
12501
MR Brain Neonatal No IVcon IMG12979
Machine Model: Agnostic
Generated:
2024-10-08 12:26:32
Use Knee Coil on Verio when possible.
1
2 *
3
4
Plane Sagittal Sagittal Axial Coronal
Sequence T1 FLASH T2 TSE T2 TSE T2 TSE
Timing --- --- --- ---
Thickness - Spacing 2-3 mm - mm 1 mm - mm 2.5 mm - mm 2.5 mm - mm
FOV 192 mm 192 mm 192 mm 192 mm
5
6 *
7 **
8 ***
9 ***
Plane Axial Axial Axial Sagittal Axial
Sequence 3D SWI T1 FLAIR DWI (B=0/500/1000) T2 HASTE T2 HASTE
Timing --- --- --- ----- -----
Thickness - Spacing 2 mm - mm 3 mm - 0.6 mm 4 mm - mm 4 mm - 1 mm 4 mm - 1 mm
FOV 192 mm 160mm 190mm approx. 200mm approx. 200mm
10 ***
sp 1
sp 2
sp 3
sp 4
Plane Coronal
sp 1
sp 2
sp 3
sp 4
Sequence T2 HASTE
sp 1
sp 2
sp 3
sp 4
Timing -----
sp 1
sp 2
sp 3
sp 4
Thickness - Spacing 4 mm - 1 mm
sp 1
sp 2
sp 3
sp 4
FOV approx. 200mm
sp 1
sp 2
sp 3
sp 4
Special Instructions:

* Add slices if need. Do not change TR
** Ensure phase enc. direction is A>P
*** Scan first if patient is moving
Indications: intraventricular hemorrhage (IVH) , neonatal hypoxic ischemic encephalopathy (HIE) , periventricular leukomalacia (PVL) , prematurity
#4
#3,5,6,7
3,5,6,7

IOML
Align series #3,5,6,7,9 parallel with subcallosal plane
If patient is 7 days old or younger, perform HIE spectroscopy (see separate protocol)
#1,8