11901
11904
CT Head No IVcon IMG14021
Machine Model: Canon Genesis 640
Generated:
2021-05-10 13:03:03
IV Contrast NA See IV Guidelines
for other access lines.
IV Saline NA
Oral Contrast NA
Water NA
Contrast amount determined by weight and renal function.
Special Instructions: IOML perpendicular to table top
Scanogram: Lateral / AP
Landmark: Hyoid
A Non-Contrast
Scan Start Mid Oral
Scan Stop Skull Vertex
Patient Position Supine
Scan Direction caudal/cranial
Scan Mode Helical
Scan Range If Volume NA
Scan Delay NA
Reconstruction 1 Thickness x Spacing .5 mm x .3 mm
Algorithm Bone
FOV 22 cm
Reconstruction 2 Thickness x Spacing 1 mm x 1 mm
Algorithm Brain
FOV 22 cm
FC 64
kVp 120 kVp
Scan FOV 22 cm
Pitch Detail
mA 350 mA
Rotation Time .75 sec
Detector Rows 0.5 mm x 40 mm
Indications: TIA / transient ischemic attack , acute bleed in the head , altered mental status , brain injury , cerebral infarct , head trauma , headache , hydrocephalus , intracranial hemorrhage , memory loss , seizures .
Age 0-6 months 6 months-3 Years 3-6 Years 6 - 12 Years 12-18 Years
CT FC 49 49 49 49 49
CT kVp 120 kVp 120 kVp 120 kVp 120 kVp 120 kVp
CT mA 250 mA 270 mA 290 mA 310 mA 350 mA
CT Rotation Time .5 sec .5 sec .5 sec .75 sec .75 sec
CTDI ----- ----- ----- ----- -----
0 images
0 images
0 images
0 images