CTV Head IVcon | IMG14312 | |
Machine Model: | Canon Aquilion ONE 320, Canon Genesis 640, Canon Vision 640 |
Generated:
2021-09-30 07:41:30
|
|
|
Special Instructions: IOML perpendicular to table top. DO NOT SEND BRAIN PERFUSION MAPS | |||||||||||
Scanogram: Lateral / AP Landmark: Hyoid | |||||||||||
A | Dynamic AOT Perfusion | A | Dynamic AOT Perfusion | A | Dynamic AOT Perfusion | A | Dynamic AOT Perfusion | A | Dynamic AOT Perfusion | ||
Scan Start | Skull Base | Skull Base | Skull Base | Skull Base | Skull Base | ||||||
Scan Stop | Skull Vertex | Skull Vertex | Skull Vertex | Skull Vertex | Skull Vertex | ||||||
Patient Position | Supine | Supine | Supine | Supine | Supine | ||||||
Scan Direction | NA | NA | NA | NA | NA | ||||||
Scan Mode | Volume | Volume | Volume | Volume | Volume | ||||||
Scan Range If Volume | 160 mm | 160 mm | 160 mm | 160 mm | 160 mm | ||||||
Scan Delay | NA | NA | NA | NA | NA | ||||||
Reconstruction 1 | Thickness x Spacing | 0.5 mm x 0.5 mm | 0.5 mm x 0.5 mm | 0.5 mm x 0.5 mm | 0.5 mm x 0.5 mm | 0.5 mm x 0.5 mm | |||||
Algorithm | Head 4D CBP | Head 4D CBP | Head 4D CBP | Head 4D CBP | Head 4D CBP | ||||||
FOV | 22 cm | 22 cm | 22 cm | 22 cm | 22 cm | ||||||
Reconstruction 2 | Thickness x Spacing | MIP 10 x 5 | MIP 10 x 5 | ||||||||
Algorithm | CTV Brain Stack | Subtraction | |||||||||
FOV | Axial + Cor + Sag | Axial + Cor + Sag | |||||||||
FC | 41 | 41 | 41 | 41 | 41 | ||||||
kVp | 80 kVp | 80 kVp | 80 kVp | 80 kVp | 80 kVp | ||||||
Scan FOV | 16 cm | 16 cm | 16 cm | 16 cm | 16 cm | ||||||
Pitch | One Shot | Intermittent | Intermittent | Intermittent | Intermittent | ||||||
mA | 500 mA | 300 mA | 430 mA | 300 mA | 300 mA | ||||||
Rotation Time | .75 sec | .75 sec | .75 sec | .75 sec | .75 sec | ||||||
Detector Rows | 0.5 mm x 160 mm | 0.5 mm x 160 mm | 0.5 mm x 160 mm | 0.5 mm x 160 mm | 0.5 mm x 160 mm | ||||||
Indications: dural sinus injury , dural sinus thrombosis . |