28554
28594
CTV Head IVcon IMG14312
Machine Model: Canon Aquilion ONE 320, Canon Genesis 640, Canon Vision 640
Generated:
2021-09-30 07:41:30
IV Contrast 5 ml/sec using 20 g PIV Minimum See IV Guidelines
for other access lines.
IV Saline 3 ml/sec using 20 g PIV Minimum
Oral Contrast NA
Water NA
Contrast amount determined by weight and renal function.
Special Instructions: IOML perpendicular to table top. DO NOT SEND BRAIN PERFUSION MAPS
Scanogram: Lateral / AP
Landmark: Hyoid
A Dynamic AOT Perfusion A Dynamic AOT Perfusion A Dynamic AOT Perfusion A Dynamic AOT Perfusion A Dynamic AOT Perfusion
Scan Start Skull Base Skull Base Skull Base Skull Base Skull Base
Scan Stop Skull Vertex Skull Vertex Skull Vertex Skull Vertex Skull Vertex
Patient Position Supine Supine Supine Supine Supine
Scan Direction NA NA NA NA NA
Scan Mode Volume Volume Volume Volume Volume
Scan Range If Volume 160 mm 160 mm 160 mm 160 mm 160 mm
Scan Delay NA NA NA NA NA
Reconstruction 1 Thickness x Spacing 0.5 mm x 0.5 mm 0.5 mm x 0.5 mm 0.5 mm x 0.5 mm 0.5 mm x 0.5 mm 0.5 mm x 0.5 mm
Algorithm Head 4D CBP Head 4D CBP Head 4D CBP Head 4D CBP Head 4D CBP
FOV 22 cm 22 cm 22 cm 22 cm 22 cm
Reconstruction 2 Thickness x Spacing MIP 10 x 5 MIP 10 x 5
Algorithm CTV Brain Stack Subtraction
FOV Axial + Cor + Sag Axial + Cor + Sag
FC 41 41 41 41 41
kVp 80 kVp 80 kVp 80 kVp 80 kVp 80 kVp
Scan FOV 16 cm 16 cm 16 cm 16 cm 16 cm
Pitch One Shot Intermittent Intermittent Intermittent Intermittent
mA 500 mA 300 mA 430 mA 300 mA 300 mA
Rotation Time .75 sec .75 sec .75 sec .75 sec .75 sec
Detector Rows 0.5 mm x 160 mm 0.5 mm x 160 mm 0.5 mm x 160 mm 0.5 mm x 160 mm 0.5 mm x 160 mm
Indications: dural sinus injury , dural sinus thrombosis .