CT Chest and Abd Pel Ped BB IVcon wo/w | IMG14236 | |
Machine Model: | Canon Aquilion ONE 320, Canon Genesis 640, Canon Vision 640 |
Generated:
2021-04-20 10:27:58
|
|
|
Special Instructions: Place and size scan over chest to include anatomy from thoracic inlet to lesser trochanter Scan coverage max: 16cm Coverage may be reduced by 2 cm increments Use WideVolume scan mode for coverage >16cm | |||||
Scanogram: AP/Lateral Landmark: Apices | |||||
A | Non-Contrast | B | Venous | ||
Scan Start | Apices | Apices | |||
Scan Stop | Lesser Trochanter | Lesser Trochanter | |||
Patient Position | Supine | Supine | |||
Scan Direction | NA | NA | |||
Scan Mode | Volume | Volume | |||
Scan Range If Volume | 160 mm | 160 mm | |||
Scan Delay | NA | 60 seconds | |||
Reconstruction 1 | Thickness x Spacing | 0.5 mm x 0.5 mm | 0.5 mm x 0.5 mm | ||
Algorithm | Ped Body | Ped Body | |||
FOV | 24 cm | 24 cm | |||
Reconstruction 2 | Thickness x Spacing | 1 mm x 1 mm | 1 mm x 1 mm | ||
Algorithm | Lung STD | Lung STD | |||
FOV | RIB to RIB | RIB to RIB | |||
FC | 18 | 18 | |||
kVp | See Chart Below Based on Weight | See Chart Below Based on Weight | |||
Scan FOV | Patient largest + 2CM | Patient largest + 2CM | |||
Pitch | NA | NA | |||
mA | modulated mA | modulated mA | |||
Rotation Time | 0.35 sec | 0.35 sec | |||
Detector Rows | 0.5 mm x 160 mm | 0.5 mm x 160 mm | |||
Indications: No indications found. . |
15KG-30KG | 0.35 Rotation Time |
31KG - 61KG+ | 0.5 Rotation Time |