MR Arthrogram Wrist Indirect w/IVcon IMG12960
Machine Model: 3T or with dedicated coil
Generated:
2023-11-17 14:57:33
Hand inject gad.; exercise affected wrist for 10 min. prior to starting scan
1
2
3
4
Plane Axial Coronal Sagittal Axial
Sequence IW FS (Intermediate-Weighted) PD FS PD FS T1
Timing Post-Contrast Post-Contrast Post-Contrast Post-Contrast
Thickness - Spacing 3 mm - 0.6 mm 2 mm - 0.4 mm 3 mm - 0.6 mm 3 mm - 0.6 mm
FOV 80-100mm 80-100mm 80-100mm 80-100mm
5
6
sp 1
sp 2
sp 3
Plane Coronal Coronal
sp 1
sp 2
sp 3
Sequence T1 FS 3D T2* FS DESS isotropic
sp 1
sp 2
sp 3
Timing Post-Contrast Post-Contrast
sp 1
sp 2
sp 3
Thickness - Spacing 2 mm - 0.4 mm 0.6 mm - mm
sp 1
sp 2
sp 3
FOV 80-100mm approx. 120mm
sp 1
sp 2
sp 3
Special Instructions:

Indications: ganglion in the wrist , internal derangement , wrist pain

If metal is in the area of interest or near enough for artifact to effect it:

- if option is available, use WARP option with VAT @ at least 50% (can go higher for larger implants)

- Use a high receiver bandwidth (approx. 500 Hz/pixel)

- for T1-weighted fat-saturated sequence, turn off all fat-saturation

- for T2-weighted fat-saturated sequences, substitute in STIR sequences with the same geometric parameters; STIR sequences must be performed pre-contrast due to the possibility of contrast-enhancing areas being nulled due to the short inversion time